Non Dispersive Infra Red Carbon Monoxide Analyzer for CO measurements in the range 0-50ppm. CO2 monitoring as option. The CO12e analyzer is an Eco-designed, reference air quality monitor, QAL1 certified by the TÜV, US-EPA approved and compliant with ISO 4224. It offers superior metrological performances for CO measurements in the range 0-300 ppm. CO2 measurement optionally (range: 0-2000 ppm by NDIR).

More about this product— External link
Most popular related searches
  • Cost-saving Carbon Monoxide monitor with ultra low power consumption offering compliance with EN 14626 (2012), EN 15267, 2008/50/EC, ISO 4224, 40 CFR PART 53 SUB B & SUB C
  • Excellent sensitivity & signal stability, great accuracy and repeatability
  • Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screens can be displayed while the instrument is operating
  • Service assistance inside: detects early signs of trouble, allows predictive maintenance, identifies the needed service and guides service operations step by step: increased productivity on site, reduced downtime, more efficiency, less training…
  • Includes Communication Protocol for XR® Software with automatic recognition and configuration
  • Automatic recognition of plugged electronic boards or optional devices: plug & play principle
  • Smart, connected instrument with embedded web server for the easiest, full remote control and operation of the analyzer from any PC, tablet or smartphone (configuration, calibration, test, diagnostic parameters…)
  • Integrated web-server with full remote emulation of the analyzer through the ENVEA Connect free app 
  • Additional parameter: CO2
  • Built-in module for CO2 monitoring (range: 0-2000 ppm by NDIR)
  • RS232 or RS485 Serial interface (via USB port)
  • External opto-isolated I/O interface with:
    • 4 independent analog inputs
    • 4 independent analog outputs
    • 4 remote control inputs
    • 6 dry contacts outputs
  • 24 V Power supply PLUS extended temperature range 0-50°C for utilization without air conditioner
  • Continuous indoor and outdoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Quality control in food and beverage industry
  • Industrial fence-line monitoring
  • Continuous emissions monitoring (CEM) by dilution
  • Medical gases: fail-safes and quality control
  • Background (urban or suburban), Rural, Traffic, Kerbside measurement campaigns and monitoring studies
  • Laboratory and field studies on CO effects
  • Measured parameter: CO (CO2 optional)
  • Technology: Gas filter Correlation – NDIR
  • US-EPA Approval / QAL 1 Certification: YES
  • Measurement: 0-50 ppm (0-300 ppm upon request)
  • Measurement units: ppb, ppm, µg/m3 (user-selectable)
  • Lower Detectable Limit (2σ): 0.05 ppm
  • Noise (σ): 0.025 ppm
  • Zero drift: < 0.1 ppm/24 hours
  • Span drift: < 1%/24 hours
  • Linearity: ± 1% FS
  • Sample flow rate: 1L/min
  • Sampling pump: Internal
  • Temperature range: +5 to +40°C (up to 50°C optionally)
  • Data storage: 1 year
  • Included I/O: 1LAN/3USB/WIFI/2DRY
  • Weight (Kg): 8
  • Dimensions mm (LxWxH): 483 x 606 x 133
  • Chassis: 19″ Rack, 3U
  • Power supply: ‘100~250Vac, 50/60Hz + ground (24V optionally)
  • Pressure and temperature compensation: YES
COMPLIANCE WITH:
  • EN 14626 (2012), EN 15267, 2008/50/EC, 40 CFR PART 53 SUB B & SUB C
  • QAL 1 certified by the TÜV to N° 0000050626
  • U.S. EPA approved as compliant with RFCA-0915-228