Hiden -Â Model HPR-20 DLS -Gas Analysis System
The Hiden QIC quartz-lined sampling interface operating at up to 200oC provides fast response times of less than 300 millisecond for most common gases and vapours, including water vapour.
The HPR-20 DLS includes Hiden’s DLS-20 triple filter quadrupole mass spectrometer with 20 mm pole diameter.
Mass range: The DLS-20 includes a unique, user switchable mode allowing the DLS-20 quadrupole mass spectrometer to operate in both the zone I and the zone H stability regions offering ultra-high resolution in the mass range to 20 amu in the stability zone H and very high resolution performance up to 200 amu in the stability zone I.
- DMM-dynamic multi-mode scanning- uniquely applied for the electron attachment ionization technique for electronegative gas studies
- Extended mass range options to 1000 amu
- Pulse ion counting electron multiplier detector with 7 decades continuous dynamic range
- QIC inlet providing
- Fast scan speeds, with up to 650 measurements per second for transient analysis
- The system is supplied with MASsoft Professional Edition software.
- This Windows 8 and Windows 10 compatible software allows control of the instrument via USB 2.0, RS232 or Ethernet. Provides comprehensive I/O for data output and for reading data from external devices for integration and display with mass spectrometer data.
Excelling in challenging application areas, the HPR-20 DLS offers outstanding sensitivity for concentration measurements of vapours and gases. Like many gas analyser manufacturers, Hiden Analytical offers the HPR-20 DLS with a choice of inlet systems for different pressure requirements.
- Mass range: 20 and 200 amu
- Minimum detectable concentration: 5 PPB
- Maximum detectable concentration: 100%
- Sampe inlet: Heated flexible silica capillary
- Standard capillary length: 2 meters
- Sample inlet pressure: 100 mbar to 2 bar abs
- Digital pulse ion counting detector: 7 decade continuous log scale
- Response time to gas concentration change: 300 milliseconds
- Fast measurement speed: Up to 650 measurements per second
- Sample flow required (Standard): 16 atm cc/min
- Sample flow required (Optional): Options for < 1 atm cc/min to > 20 atm cc/min
- Corrosive gas sampling options: Yes