PlasmoTest is the first cellular assay, designed as a routine method for the visual, colorimetric detection of mycoplasma contamination in cell cultures.

Most popular related searches

Mycoplasma contamination of cultured cells is a major problem in scientific research. It cannot be detected by visual inspection and may remain totally undetected for long periods, thus affecting virtually any function of eukaryotic cells and leading to unreliable experimental results and potentially unsafe biological products.

Thus, InvivoGen introduces PlasmoTest™, a cell-based, reliable, and sustainable tool to detect mycoplasma contamination in cell culture. It features two major components:

  • the endotoxin sensor cells, HEK-Blue™-2 cells
  • HEK-Blue™ Detection medium

Assay principle

HEK-Blue™-2 cells, engineered to become extremely sensitive, recognize even small amounts of mycoplasma through the Toll-like receptor (TLR) 2. This leads to the activation of NF-κB and the subsequent release of SEAP, a secreted embryonic alkaline phosphatase reporter protein, in the supernatant.

These cells are cultured in HEK-Blue™ Detection medium. In the presence of SEAP, it changes its color to purple/blue, making it ideal for routine tests. The simple addition of test samples to these cells provides colorimetric results with sensitivity similar to luminescence-based biochemical assays. SEAP activity can also be measured using a spectrophotometer for a more precise and semi-quantitative result.

  • Simple - Requires basic cell culture knowledge and no specific lab equipment
  • Quick - Hands-on time < 1 hour. Results after overnight incubation
  • Reliable- No false positive: a positive result indicates the presence of a cell culture contaminant
  • Versatile - Indicates mycoplasma and bacterial contamination
  • Highly sensitive - Detects 5.102-5.105 cfu/ml mycoplasma contaminants
  • Economical - 1 kit allows to perform up to 250 tests

PlasmoTest™ cannot be used for Mycoplasma species detection. However, it can detect other cell culture contaminants, such as bacteria.

  • APPLICATIONS: Mycoplasma and bacterial detection in cell cultures
  • LIMIT OF DETECTION: 5.102- 5.105 cfu/ml (depending on the species of mycoplasma)
  • DETECTION METHOD: Colorimetric
  • CELL TYPE: Epithelial
  • GROWTH PROPERTIES: Adherent
  • TISSUE ORIGIN: Human embryonic kidney cells
  • GROWTH MEDIUM: Complete DMEM (see TDS)
  • TESTED APPLICATIONS: Tested on the five mycoplasma species most frequently isolated from cell cultures
  • QUALITY CONTROL: Each lot is functionally tested and validated.
  • ADDITIONAL INFORMATION: No false positive: a positive result indicates the presence of a cell culture contaminant