Prozess Analysen Instrumente (PAI)

Model FL200 -On-line Fluorometer

SHARE

Mixed use: portable or fixed installations. Built-in sampling pump and battery. Measurement within 5 seconds. Low maintenance and operating costs. Compact size.

Most popular related searches

Based on UV or visible fluorescence, a wide range of applications can be covered by the FL200 depending on the optical head used:

  • FL200-C: Chlorophyll A
  • FL200-H: PAH (aromatic hydrocarbons) in water (BTEX, Phenol, oil, fuel)
  • FL200-R: Rhodamine
  • FL200-F: Fluorescein

Using a high sensitivity photomultiplier, the FL200 can measure very low concentrations in the ppb range.

Main applications

Chlorophyll A

  • River, lake, reservoir surveys
  • Drinking water treatment plants
  • fish farms

PAH

  • River surveys
  • Ground water control cleansing
  • Drinking water treatment plants
  • Cooling water survey

Fluorescent tracers

  • Ground water contaminant studies for chemical plants
  • Flow calibration of water treatment plants
  • Flow measurement in streams.

Very low operating costs

  • The UV spectroscopy measuring principle requires no chemical reagent or calibration solutions resulting in very low operating and maintenance costs.

Automatic cleaning system

  • A fully automated cleaning system prevents the measurement flow cell from becoming dirty, giving the analyser autonomy for several weeks without maintenance.
  • The cleaning solution (5% sulphuric acid) should be renewed once a month.


Built-in peristaltic pump

  • When the water is not pressurised (rivers, effluents, sewage), a peristaltic pump can be added to the analyser. It is synchronized with the measurements to increase the lifetime of the tubes.

Battery/mains power supply

  • For field measurements or isolated sites, a 12V built-in battery can make the analyser autonomous for about 100 measurements.
  • For plant applications, the battery provides total immunity against mains disturbances or power cuts, even over a long period.

Built-in datalogger

  • The measurements are dated and stored in a static memory with a capacity of more than 10,000 measurements.
  • They can be tranferred later via the RS232 port on a PC without specific software using
  • Hyperterminalâ of Windowsâ . The data are compatible with standard worksheets, particularly Excelâ , to obtain graphs easily.

Graphic display

  • Measurements can be displayed on the graphic screen showing all data obtained during one hour, one day, one week, one month or one year.
  • During the measurement cycle, a moving synoptic shows the operation sequence.

Options

  • Peristaltic sampling pump
  • Measurement remote command
  • Modem board

Specifications on-line fluorometer FL200

  • Range: FL200-C: 0 - 300 µg/l Chl A
  • FL200-H: 0 - 10 mg/l phenol equiv.
  • FL200-R: 0 - 1000 ppb Rhodamine
  • FL200-F: 0 - 1000 ppb Fluorescein
  • Repeatability FL200-C: +/- 0.3 µg/l Chl. A typical
  • Repeatability FL200-H: +/- 3 µg/l Phenol equiv. typical
  • Repeatability FL200-R: +/- 0.03 ppb Rhodamine typical
  • Repeatability FL200-F: +/- 0.03 ppb Fluorescein typical
  • Initial calibration: +/- 2% typical
  • Sample input/output: Stainless steel fitting for plastic tube external diameter 12 mm
  • Pressure: Maximum 5 Bar
  • Flow: 0 - 5 L/mn, typical 0.5 L/mn
  • Sample temperature: > 0°C - 50°C
  • Outputs: 4-20 mA insulated,12 bit resolution
  • High and low threshold relays
  • Communication: Port 1: RS232 for PC or modem or MODBUS
  • Port 2: RS232 for on-line printer
  • Power supply: 110-120 / 220-240V 50/60 Hz 30VA
  • + built-in 12V battery
  • Casing: Watertight IP559 - Ambient temperature > 0°C - 60°C
  • Weight: 13 Kg /18 Kg without/with cleaning system
  • Standards: CE Conformity - EN50081-2, EN50082-2, EN55011