I.S.E. Srl (ISE Srl)

Instruments Maintenance Services

SHARE

Replace all peristaltic pump heads (about 600,000 machine working cycles in 1 year). Replace photometer halogen lamp for optimized performances (2,000 hours or 1 year max) and then run a Gain Calibration cycle. Replace the sampler probe/s for optimized performances (replace earlier if damaged). Replace tanks filters with new ones for best operation. Replace the in-line filter of the washing sink draining diaphragm pump inlet (P3140000324) with a new one to avoid obstruction (not present in Miura One and Hemo One). Clean and disinfect the instrument probe washing sink to remove soils. Verify instrument operation and check positions coordinates. Replace the washing station white tip on the last position.